To assess dosimetric parameters in a case study where bilateral accelerated partial breast irradiation (APBI) is delivered using a strut-adjusted volume implant (SAVI) device. A 59-year-old female received APBI in both breasts over 5 days, with fractions of 3.4 Gy twice daily. A Vac-lok system was used for immobilization, and a C-arm was used for daily imaging. We generated dose-volume histograms (DVHs) for the brachytherapy plans to derive several important biologic factors. We calculated the normal tissue complication probability (NTCP), equivalent uniform dose (EUD), and tumor control probability (TCP) using the Lyman-Kutcher-Burman model parameters α = 0.3 Gy(-1), α/β = 4 Gy, n = 0.1, and m = 0.3. In addition, we assessed the dose homogeneity index (DHI), overdose index, and dose nonuniformity ratio. D95 was >95