We have developed a simple and sensitive method for the simultaneous determination of testosterone (TES), cortisol (CRT), and dehydroepiandrosterone (DHEA) in saliva by automated online in-tube solid-phase microextraction (SPME) coupled with liquid chromatography-tandem mass spectrometry (LC-MS/MS) using a Discovery HS F5 column. The optimum in-tube SPME conditions were 25 draw/eject cycles of 40 μL of sample at a flow rate of 200 μL/min using a Supel-Q PLOT capillary column as an extraction device. The extracted compounds were easily desorbed from the capillary by passage of the mobile phase, and no carryover was observed. The in-tube SPME LC-MS/MS method showed good linearity with correlation coefficients r ≥ 0.9998 for TES, CRT, and DHEA using their respective stable isotope-labeled internal standards. The intra-day and inter-day precisions (relative standard deviations) were below 4.9 and 8.5
作者:Hiroyuki, Kataoka;Kentaro, Ehara;Rie, Yasuhara;Keita, Saito
来源:Analytical and bioanalytical chemistry 2013 年 405卷 1期