This case study compares the impact of two prosthetic socket designs, a "traditional" transhumeral (TH) socket design and a Compression Released Stabilized (CRS) socket.A CRS socket was compared to the existing socket of two persons with transhumeral amputation. Comparisons included assessments of patient comfort and satisfaction with fit, as well as dynamic kinematic assessment using a novel high-speed, high-resolution, bi-plane video radiography system (XROMM, for X-ray Reconstruction of Moving Morphology).Subjects were more satisfied with the comfort of the traditional sockets, although they had positive impressions about aspects of the fit and style of the CRS socket, and thought that it provided better control. Dynamic kinematic assessment revealed that the CRS socket provided better control of the residual limb within the socket, and had less slippage as compared to a traditional TH socket design.The TH CRS socket provided better control of the residual limb within the socket, and had less slippage. However, participants were less satisfied with the comfort and overall utility of the CRS socket, and stated that additional fitting visits/modifications to the CRS socket were needed. It is possible that satisfaction with the CRS socket may have improved with prosthetic adjustment and more acclimation time. Implications for Rehabilitation A comfortable, good fitting prosthetic socket is the key factor in determining how long (or if) an upper limb amputee can tolerate wearing a prosthesis. This case series was a comparison of two socket designs, a 'traditional' socket design and a Compression Released Stabilized (CRS) socket design in persons with transhumeral amputation. The CRS socket provided better control of the residual limb within the socket, and had less slippage. However, its tightness made it more difficult to don. Both subjects were less satisfied with the comfort and overall utility of the CRS socket. However, satisfaction might have been improved with additional fitting visits and more acclimation time.

作者:Linda, Resnik;Tarpit, Patel;Shane G, Cooney;Joseph J, Crisco;Christopher, Fantini

来源:Disability and rehabilitation. Assistive technology 2016 年 11卷 5期

相似文献
知识库介绍

临床诊疗知识库该平台旨在解决临床医护人员在学习、工作中对医学信息的需求,方便快速、便捷的获取实用的医学信息,辅助临床决策参考。该库包含疾病、药品、检查、指南规范、病例文献及循证文献等多种丰富权威的临床资源。

详细介绍
热门关注
免责声明:本知识库提供的有关内容等信息仅供学习参考,不代替医生的诊断和医嘱。

收藏
| 浏览:37
作者:
Linda, Resnik;Tarpit, Patel;Shane G, Cooney;Joseph J, Crisco;Christopher, Fantini
来源:
Disability and rehabilitation. Assistive technology 2016 年 11卷 5期
标签:
Amputation biomechanics evaluation kinematic motion prosthesis radiography upper limb
This case study compares the impact of two prosthetic socket designs, a "traditional" transhumeral (TH) socket design and a Compression Released Stabilized (CRS) socket.A CRS socket was compared to the existing socket of two persons with transhumeral amputation. Comparisons included assessments of patient comfort and satisfaction with fit, as well as dynamic kinematic assessment using a novel high-speed, high-resolution, bi-plane video radiography system (XROMM, for X-ray Reconstruction of Moving Morphology).Subjects were more satisfied with the comfort of the traditional sockets, although they had positive impressions about aspects of the fit and style of the CRS socket, and thought that it provided better control. Dynamic kinematic assessment revealed that the CRS socket provided better control of the residual limb within the socket, and had less slippage as compared to a traditional TH socket design.The TH CRS socket provided better control of the residual limb within the socket, and had less slippage. However, participants were less satisfied with the comfort and overall utility of the CRS socket, and stated that additional fitting visits/modifications to the CRS socket were needed. It is possible that satisfaction with the CRS socket may have improved with prosthetic adjustment and more acclimation time. Implications for Rehabilitation A comfortable, good fitting prosthetic socket is the key factor in determining how long (or if) an upper limb amputee can tolerate wearing a prosthesis. This case series was a comparison of two socket designs, a 'traditional' socket design and a Compression Released Stabilized (CRS) socket design in persons with transhumeral amputation. The CRS socket provided better control of the residual limb within the socket, and had less slippage. However, its tightness made it more difficult to don. Both subjects were less satisfied with the comfort and overall utility of the CRS socket. However, satisfaction might have been improved with additional fitting visits and more acclimation time.